Accuracy is one of the most important characteristics of a load cell.
Accurate information is essential not only when weighing, but also in all the process management and control operations.
But what is to be done if there is a fault or malfunction on the weighing system and the load cell measurement is no longer so accurate?
There are 2 ways to check that a load cell is working properly: through built-in diagnostics offered by some instruments and through a digital multimeter test.
This is the on-board function of some multichannel weight transmitters and some junction boxes, such as the TLM8 and TLB4 weight transmitters, or the intelligent junction box CLM8.
If the load cell is connected to an instrument offering this possibility, the malfunction is reported directly by the weight transmitter or the junction box.
→ Read also "How to check if a load cell is working correctly - Built-in diagnostics".
This is an in-depth test that is helpful when you want to measure both the resistance and the no-load voltage values of the load cell.
Here's a detailed look at how to perform the two tests.
1. The first thing to do is to disconnect the load cells from the instrument and check that there are no traces of moisture in the junction box due to water ingress or condensation.
If there are any, the system must be drained or, if necessary, replaced.
2. Connect the digital multimeter to the load cell you want to test.
3. Using the multimeter, check:
To measure the voltage, the load cell must be tested unloaded, i.e. with no load applied.
Simply remove it from the weighing system or, if the mounting kit has one, the devices can be activated to completely unload the load cell.
This possibility is provided, for example, by the LAUMAS V10000 and VCOKDTL mounting kits.
Once the load cell has been unloaded, the following steps must be carried out: