تحرير الملاحظات

Ver. 1.18.01 (18 Dec 2024)

Added

  • Modbus protocol: if coefficient is active, gross and net weight, with the coefficient applied, are published

Fixed

  • CANOPEN: MODBUS communication through RS485 port returns several errors during data exchange

Ver. 1.18.00 (25 Nov 2024)

Added

  • Firmware version compliant with INMETRO regulation
  • Modbus protocol: address 0 supported for broadcast communication
  • Analog output: output current or voltage value configured by remote
  • RS485 serial port: continuous transmission YHL protocol available

Ver. 1.17.00 (24 Oct 2023)

Fixed

  • When a semi-automatic tare is active, gross weight temporary visualization, pressing Up arrow key, doesn't work

Ver. 1.16.05 (22 May 2023)

Fixed

  • Modified one of the voltage thresholds used by diagnostic functions, to avoid load cell error condition incorrect signalings

Ver. 1.16.04 (18 Apr 2023)

Fixed

  • The reset procedure was changed because after resetting the instrument to the factory defaults, the weight was not calculated correctly.

Ver. 1.16.03 (02 Dec 2022)

Fixed

  • Threshold for load cell error alarm modified, to avoid wrong signallings

Ver. 1.16.02 (17 Nov 2022)

Fixed

  • ASCII protocol: fixed semiautomatic zero and semiautomatic tare commands working

Ver. 1.16.01 (19 Jul 2022)

‫Internal changes without any impact on the user

Ver. 1.16.00 (26 Jan 2022)

‫Internal changes without any impact on the user

Ver. 1.15.02 (21 Oct 2020)

Fixed

  • Communication errors, in case of MODBUS reading if more than 10 registers are required

Ver. 1.15.00 (22 Sep 2020)

Added

  • Real calibration, command code 106: a new calibration point is added
  • Real calibration, checks executed when a new calibration point is defined:
    • Operation not executed if the sample weight is equal to 0
    • Operation not executed if the sample weight is used in one of the currently defined calibration points

Changed

  • Real calibration, command code 101 original functioning restored. It deletes all previously registered calibration points and add a new one

Ver. 1.14.16 (18 Jun 2020)

Changed:

  • Relay outputs maintained open during instrument startup

Ver. 1.14.13 (11 Jan 2020)

‫Internal changes without any impact on the user

Ver. 1.14.11 (29 Nov 2019)

Added

  • First firmware release compliant with Instrument Manager service software

Ver. 1.12.07 (03 Oct 2019)

Fixed

  • Analog output, 0-20 mA o 4-20 mA types: minimum output current value set to -0.5 mA, maximum output current value set to 22 mA
  • Analog output, 0-5 V 0 0-10 V types: minimum output voltage value set to -0.5 V, maximum output voltage value set to 10.2 V 

Ver. 1.12.06 (11 Oct 2018)

Fixed

  • Wrong syntax for transmiteed string, in case of CONTIN protocol, MOD TD mode active

Ver. 1.12.05 (29 May 2018)

‫Internal changes without any impact on the user

Ver. 1.12.03 (21 May 2015)

Fixed

  • If RIP, HDRIP or HDRIPN protocol enabled, peak weight transmitted, instead of the net weight, when peak function is active

Ver. 1.12.01 (17 Jul 2014)

Added

  • Specific menu for anti peak filter setup

Ver. 1.11.02 (19 Jan 2012)

Fixed

  • Setpoint menu, Setp1 label changed in Set 01. The other setpoint labels has been changed in the same way

Ver. 1.11.00 (02 Dec 2011)

Added:

  • MODBUS-RTU: preset tare execution