Ver. 1.12.00 (15 Jan 2025)
Added
- Firmware version compliant with INMETRO regulation
- Modbus protocol: address 0 supported for broadcast communication
- Modbus protocol: if coefficient is active, gross and net weight, with the coefficient applied, are published
|
Ver. 1.11.10 (02 Jul 2024)
Fixed:
- Low battery condition signalled with delay
|
Ver. 1.11.08 (14 Jun 2024)
Fixed:
- Zero setting and alibi memory are kept after update and after low battery buffer error
|
Ver. 1.11.06 (15 Jan 2024)
Internal changes without any impact on the user |
Ver. 1.11.04 (27 Dec 2023)
Fixed
- Batching start with external contact. instrument freezed and display showing message 'F X'
|
Ver. 1.11.00 (01 Dec 2023)
Added:
- Label printer management, in case of Zebra models or printers supporting ZPL programming language
- Semiautomatic or predetermined tare maintained at the start of a batching sequence
- Full management of batching functions allowed in case of Profinet-IO, Ethernet/IP, CANOpen and DeviceNET communication interfaces
Fixed:
- Legal for trade mode, NTEP: Pounds available as unit of measure
Modified:
- The following functions, regarding batching programs, can be freely activated by the user:
- Intermediate unloadings
- Cycle end partial unloadings
- Setting of a quantity to be batched greater than the scale capacity
- Setting formulas in percentages
- Displaying of batching data on remote displays
|
Ver. 1.10.20 (08 Jun 2023)
Fixed
Preset tare: corrected mV/weight calculation with piece conuter mode. Corrected field bus setting |
Ver. 1.10.16 (09 Mar 2023)
Fixed
- Semi-automatic zero from input doesn't control alarm conditions
|
Ver. 1.10.14 (15 Nov 2022)
Fixed
- Device's programs wasn't selected correctly
|
Ver. 1.10.12 (11 Nov 2022)
Fixed
- Batching programs wrongly recognized by Instrument Manager
|
Ver. 1.10.10 (10 Nov 2022)
Fixed
Added error message in case of USB port malfunction ("USB KO") |
Ver. 1.10.08 (11 Oct 2022)
Fixed
- Correct quick access to the tare weight zero setting
|
Ver. 1.10.04 (04 Aug 2022)
Fixed
- Corrections when operating with intelligent junction box
- Corrections to real calibration via Instrument Manager
- Corrections to ASCII protocol
|
Ver. 1.10.04 (04 Aug 2022)
Fixed
- Corrections when operating with intelligent junction box
- Corrections to real calibration via Instrument Manager
- Corrections to ASCII protocol
|
Ver. 1.10.02 (25 Mar 2022)
Fixed
- ASCII protocol: NET command lights up the corresponding LED but the weight is not zeroed
|
Ver. 1.10.00 (17 Feb 2022)
Added
- NTEP compliant version
- Add EPSON TMT20III printer type
Fixed
- Removed restriction on the voltage/current value produced by the analog output related to calculated weight
|
Ver. 1.09.06 (13 Jul 2021)
Fixed
- Setting weight to zero by input 1 doesn't show error message "t-----"
|
Ver. 1.09.05 (11 Jun 2021)
Fixed
- Fix parameters GATEWAY, SUBNET MASK managements for OPZW1RADIO
|
Ver. 1.09.04 (20 Oct 2020)
Fixed
- Update firmware via Instrument Manager: Set program and options before erase step. Fix menu items visualization.
- INFO menu: active options wrong visualization
|
Ver. 1.09.02 (29 Sep 2020)
Internal changes without any impact on the user |
Ver. 1.09.00 (21 Sep 2020)
Added
- First release Instrument Manager compliant (Version 1.1.0)
- Remote real calibration up to 8 points, added by command 106. Values can be negative.
- PEAK input function run via the START/STOP key
|
Ver. 1.08.24 (22 Jun 2020)
Fixed
- Continuous weight transmission: doesn't display NO COM message when serial communication doesn't work
|
Ver. 1.08.22 (23 Jan 2020)
Added
- Connected to junction box: Add tilt alarm visualization (tilt, comm, maxx, maxy)
Fixed
- Connected to junction box: weight values aren't refreshed after change division
|
Ver. 1.08.20 (18 Dec 2019)
FIXED
- UNLOAD: fix SET closing contact during batching after enter in the menus
|
Ver. 1.08.18 (20 May 2019)
Fixed
- Setting P1 turn on input led 1
|
Ver. 1.08.16 (02 Apr 2019)
Fixed
- ASCII: Commands 'z' and 's' doesn't reply 0 and setted sample weight
- HDRIP protocol: reply gross weight instead net weight
- Wrong visibility on approved program of IN3 and EC with analog option
|
Ver. 1.07.70 (08 Feb 2019)
Internal changes without any impact on the user |
Ver. 1.07.68 (19 Jan 2019)
Internal changes without any impact on the user |
Ver. 1.07.66 (05 Jan 2019)
Internal changes without any impact on the user |
Ver. 1.08.14 (22 Dec 2018)
Fixed
Analog Output: not save all parameters when operating mode is changed (from current to tension). At start up the instrument does not work properly. |
Ver. 1.08.12 (07 Dec 2018)
Fixed
- UNLOAD program: no display of ALARM menu
|
Ver. 1.08.10 (09 Nov 2018)
Added
- Peak weight management from button
- Unlock menu with password
- Add input LIMIT SWITCH mode
- NET EN mode: if YES, disable tare functions on approved program instrument
- WEIMOD: add on approved program instrument
|
Ver. 1.07.64 (17 Oct 2018)
Added
CONTIN: add functionality for TCP-IP on serial 3 |
Ver. 1.07.42 (23 Aug 2018)
Internal changes without any impact on the user |
Ver. 1.07.40 (18 Jul 2018)
Internal changes without any impact on the user |
Ver. 1.07.62 (17 Jul 2018)
Added
- With OPZW1RADIO is possible to manage instrument TLKWF as junction box (MASMUL mode on series W serial communication)
Correzioni
- Access to legally relevants parameters lost at switch off
- IMP0: the value is set only if the actual value is changed
- ALIFUL message: can be resetted by Modbus
- PARSER alarm (Batching): pressing ENTER all formulas are rewritten automatically
- FS-TEO: Parameter can't be changed
- USB option: managed diffetent USB model types
- Autotare delay and Stable tare must be visible only in LOAD programs
- By displaying gross weight while net weigth is set, NET/GROSS led does not update correctly
- In case of multiple print with access to legal parameter instrument the alibi ID was incremented in the print copies
- Fix tare rounding with access to legal parameters instrument
- Intelligent junction box: even without communication, legal parameter access is managed with password on the transmitter
- Intelligent junction box: fixed display of negative percentages in diagnostics
- Intelligent junction box: fixed peak weight management
- Intelligent junction box: print diagnostic event does not print percentages
- Intelligent junction box: fixed filter A visualization
|
Ver. 1.07.22 (16 Apr 2018)
Internal changes without any impact on the user |
Ver. 1.07.20 (23 Feb 2018)
Fixed
- After set WIFI option, in option menu is shown NONE instead of WIFI
- With WIFI option on, in INFO menu is shown NO OPZ message
|
Ver. 1.07.10 (16 Nov 2016)
Added
- WDOS TOTAL: added continuos weight trasmission protocol. Repeater instrument shows TOTAL weight.
|
Ver. 1.07.08 (29 Sep 2016)
Fixed
- Modbus-TCP/IP: wrong menu visibility
|
Ver. 1.07.06 (28 Sep 2016)
Fixed
- Filter non visible in batching programs
|
Ver. 1.07.04 (23 Aug 2016)
Fixed
- Updated opzw1radio management
- Profinet-IO: Real calibration works in a system composed by an indicator and an intelligent junction box
|
Ver. 1.06.32 (13 Jul 2016)
Internal changes without any impact on the user |
Ver. 1.06.30 (17 Jun 2016)
Fixed
- Research in progress (Alibi) bit in Modbus status register was not refreshed
|
Ver. 1.06.28 (05 May 2016)
Internal changes without any impact on the user |
Ver. 1.06.26 (22 Mar 2016)
Fixed
- Contrast value visualization
|
Ver. 1.06.24 (17 Mar 2016)
Internal changes without any impact on the user |
Ver. 1.06.22 (11 Feb 2016)
Fixed
- No comparison time: load batching program not correctly managed
|
Ver. 1.06.20 (17 Jun 2015)
Fixed
- Preset Tare applied for pieces count
- Profinet-IO: change store parameters in eeprom. Legal for trade instrument: instrument working extremely low
|
Ver. 1.06.16 (18 Mar 2015)
Added
- Weigth input selectable from serial options
|
Ver. 1.06.14 (09 Feb 2015)
Fixed
- RIPLAU protocol blocks the instrument
|
Ver. 1.06.10 (23 Jun 2014)
Added
- Profinet IO: Parameters IP ADDR, GATEWAY and MASK can be write and store from fieldbus
Fixed
- Outputs management
- Inputs management on TOTAL Program
- Contin Protocol on TCP/IP
- Setpoints reading on Modbus
|